Loading...
Please wait, while we are loading the content...
Light-Scattering Method in Particle Size Analysis of Parenteral Emulsions
| Content Provider | Scilit |
|---|---|
| Author | Tian, Youqin Chiu, Luk |
| Copyright Year | 1998 |
| Description | Journal: Drug Development and Industrial Pharmacy The use of a light-scattering particle size distribution analyzer has been shown to be a convenient method for characterizing the particle size distribution of parenteral emulsions. However, the concentrations of the samples used were found to have a major impact on the particle size distribution results, particularly for samples with a mean particle size smaller than 0.2 micron. An increase in sample concentration caused a shift to smaller particle sizes as a result of multiple scattering. The blue-light (tungsten lamp) intensity, instead of the He-Ne laser integrity, should be used to control sample concentration within the optimal range for measurement. |
| Related Links | http://www.tandfonline.com/doi/pdf/10.3109/03639049809085620 |
| Ending Page | 280 |
| Page Count | 6 |
| Starting Page | 275 |
| ISSN | 03639045 |
| e-ISSN | 15205762 |
| DOI | 10.3109/03639049809085620 |
| Journal | Drug Development and Industrial Pharmacy |
| Issue Number | 3 |
| Volume Number | 24 |
| Language | English |
| Publisher | Informa UK Limited |
| Publisher Date | 1998-01-01 |
| Access Restriction | Open |
| Subject Keyword | Journal: Drug Development and Industrial Pharmacy Pharmacology and Pharmacy Light Scattering Parenteral Emulsions Particle Size |
| Content Type | Text |
| Subject | Organic Chemistry Drug Discovery Pharmacology Pharmaceutical Science |