Recent Progress in Birdcage RF Coil Technology for MRI System
| Content Provider | MDPI |
|---|---|
| Author | Ahmad, Sheikh Faisal Kim, Young Cheol Choi, Ick Chang Kim, Hyun Deok |
| Copyright Year | 2020 |
| Description | The radio frequency (RF) coil is one of the key components of the magnetic resonance imaging (MRI) system. It has a significant impact on the performance of the nuclear magnetic resonance (NMR) detection. Among numerous practical designs of RF coils for NMR imaging, the birdcage RF coil is the most popular choice from low field to ultra-high field MRI systems. In the transmission mode, it can establish a strong and homogeneous transverse magnetic field B1 for any element at its Larmor frequency. Similarly, in the reception mode, it exhibits extremely high sensitivity for the detection of even faint NMR signals from the volume of interest. Despite the sophisticated 3D structure of the birdcage coil, the developments in the design, analysis, and implementation technologies during the past decade have rendered the development of the birdcage coils quite reasonable. This article provides a detailed review of the recent progress in the birdcage RF coil technology for the MRI system. |
| Starting Page | 1017 |
| e-ISSN | 20754418 |
| DOI | 10.3390/diagnostics10121017 |
| Journal | Diagnostics |
| Issue Number | 12 |
| Volume Number | 10 |
| Language | English |
| Publisher | MDPI |
| Publisher Date | 2020-11-27 |
| Access Restriction | Open |
| Subject Keyword | Diagnostics Nuclear Energy and Engineering Birdcage Coil Rf Coil Volume Coil Numerical Method Analytical Solution Electromagnetic Simulation Mri Nmr Imaging Design Issues Dual Resonance |
| Content Type | Text |
| Resource Type | Article |