Loading...
Please wait, while we are loading the content...
Similar Documents
A technique for dynamically calibrating pressure transducers at cryogenic temperatures
| Content Provider | NASA Technical Reports Server (NTRS) |
|---|---|
| Author | Gibbens, B. V. |
| Copyright Year | 1983 |
| Description | A technique was developed for the calibration of dynamic pressure transducers at cryogenic temperatures. The calibration system utilizes an 8.9 Newton peak thrust shaker which oscillates a helium-filled bellows to generate a sinusoidal dynamic pressure to calibrate transducers immersed in a cryogenic environment. The system has a dynamic pressure measurement uncertainty of approximately 11% and is capable of producing peak-to-peak dynamic pressure amplitudes of 1.4 kPa over a frequency range of 40 to 100 hertz and a temperature range of 100 to 300 K. It provides an unprecedented capability of both static and dynamic calibration of pressure transducers from ambient to cryogenic temperature. |
| File Size | 574078 |
| Page Count | 17 |
| File Format | |
| Alternate Webpage(s) | http://archive.org/details/NASA_NTRS_Archive_19840003390 |
| Archival Resource Key | ark:/13960/t16m82z9w |
| Language | English |
| Publisher Date | 1983-10-01 |
| Access Restriction | Open |
| Subject Keyword | Pressure Sensors Bellows Shakers Error Analysis Calibrating Cryogenics Pressure Measurement Cryogenic Temperature Displacement Dynamic Pressure Ntrs Nasa Technical Reports ServerĀ (ntrs) Nasa Technical Reports Server Aerodynamics Aircraft Aerospace Engineering Aerospace Aeronautic Space Science |
| Content Type | Text |
| Resource Type | Technical Report |