Precise efficiency calibration of an HPGe detector: source measurements and Monte Carlo calculations with sub-percent precision.
| Content Provider | Semantic Scholar |
|---|---|
| Author | Hardy, Joanne Iacob, Viorica Sánchez-Vega, M. Gabriela Effinger, Robert T. Lipnik, Pascale Mayes, Veronica Willis, David Kyle Helmer, Richard |
| Copyright Year | 2002 |
| Abstract | With the goal of measuring precise gamma-ray intensities for short-lived (< 5 s) accelerator-produced activities, we have calibrated the efficiency of an HPGe detector between 53 and 1836keV to sub-percent precision with a combination of source measurements and Monte Carlo calculations. Using known or independently measured detector dimensions, we have achieved both relative and absolute agreement (the latter, to 0.1%) between the calculated and measured efficiencies with only two adjustable detector parameters, the thicknesses of the contact dead layers. |
| Starting Page | 1 |
| Ending Page | 3 |
| Page Count | 3 |
| File Format | PDF HTM / HTML |
| Alternate Webpage(s) | https://cyclotron.tamu.edu/progress-reports/2000-2001/5_a10x.pdf |
| PubMed reference number | 11839060v1 |
| Volume Number | 56 |
| Issue Number | 1-2 |
| Journal | Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine |
| Language | English |
| Access Restriction | Open |
| Subject Keyword | Calibration Detectors Dimensions Efficiency Monte Carlo Method anatomical layer |
| Content Type | Text |
| Resource Type | Article |