| Abstract | RESULTS When AAA 1-mm CGS plans and AAA 2.5-mm CGS plans were compared, 4% difference was observed; when AXB 1-mm CGS plans and AXB 2.5-mm CGS plans were compared, 1% difference was found. No significant difference was found between plans with AAA 1-mm CGS and plans with AXB 1-mm CGS (p>0.05). On the other hand, there was a significant difference between plans with AAA 2.5mm CGS and plans with AXB 2.5-mm CGS (p<0.05). |