Single wavelength check isocratic reverse phase HPLC method for fast and direct estimation of alkyl aromatic oxidation products
| Content Provider | Semantic Scholar |
|---|---|
| Author | Munshi, Pradip Pillai, S. Muthukumaru Rao, Gandham Satya Srinivasa |
| Copyright Year | 2010 |
| Abstract | An isocratic, reverse phase liquid chromatographic method has been developed for the quantitative determination of p-xylene oxidation products. The method represents a direct and fast analytical procedure to determine the assay. Separation is achieved on Waters μ-Bondapak C18, 300 mm × 3.9 mm × 10 μm column in a single wavelength 230 nm. The method has been validated with respect to linearity, limit of quantification, accuracy and robustness. Critical impurity 4-carboxy benzaldehyde (4-CBA) is found to be within the limit of quantification, 8 ppm, accuracy 97–98%, recovery 98 to 101% and resolution 1.97 from the closest peak. The sample solution is stable and can be used safely within 48 h. The method is verified by real sample reaction mixture. |
| Starting Page | 382 |
| Ending Page | 386 |
| Page Count | 5 |
| File Format | PDF HTM / HTML |
| DOI | 10.1039/B9AY00081J |
| Volume Number | 2 |
| Alternate Webpage(s) | http://lib3.dss.go.th/fulltext/Journal/analytical%20method/2010%20v2n4/2010%20v2n4%20p382-386.pdf |
| Alternate Webpage(s) | https://doi.org/10.1039/B9AY00081J |
| Language | English |
| Access Restriction | Open |
| Content Type | Text |
| Resource Type | Article |