Interactive comment on “Underway seawater and atmospheric measurements of volatile organic compounds in the Southern Ocean” by Charel Wohl et al
| Content Provider | Semantic Scholar |
|---|---|
| Author | Referee, Byron Blomquist |
| Copyright Year | 2020 |
| Abstract | I wonder about sampling artifacts in the atmospheric measurements using the long inlet, described only as 90 meter, 9.5mm OD teflon tube. This inlet is probably OK for DMS but may contribute measurement bias for the more soluble species. Can the authors cite evidence to address this concern? Diverting sample flow to a Pt-catalyst combustion furnace provides an instrument blank for the PTRMS, but the authors don’t mention doing a zero-air or standard injection at the inlet tip to characterize sampling |