Loading...
Please wait, while we are loading the content...
Similar Documents
Comparison of valve performance of the mechanically expanding Lotus and the balloon-expanded SAPIEN3 transcatheter heart valves: an observational study with independent core laboratory analysis.
| Content Provider | Semantic Scholar |
|---|---|
| Author | Soliman, Osama I. I. Faquir, Nahid El Ren, Ben Spitzer, Ernest Gils, Lennart Van Jonker, Hans A. J. Geleijnse, Marcel L. Es, Gerrit-Anne Van Tijssen, Jann G. P. Mieghem, Nicolas Mda Van Jaegere, Peter P. T. De |
| Copyright Year | 2018 |
| Abstract | Aims The Lotus and SAPIEN3 are second-generation transcatheter heart valves, which are designed to minimize paravalvular aortic regurgitation (PAR) after transcatheter aortic valve replacement (TAVR). We sought to compare both devices for valve performance and with emphasis on PAR by independent core laboratory analysis. Methods and results A total of 162 (79 Lotus and 83 SAPIEN3) consecutive patients (51% female, 80 ± 8 years, Logistic EuroSCORE 14.8 ± 9.4%) who underwent TAVR because of aortic stenosis were included. Patients with aortic valve-in-valve treatment were excluded. Pre-discharge echocardiograms were analysed by an independent core laboratory using the Valve Academic Research Consortium 2 criteria. There were no differences in baseline and procedural characteristics, except for a larger aortic annulus and sizing indices in SAPIEN3-treated patients and frequency of post-dilatation (0% in Lotus and 13.1% in SAPIEN3). Both valves have similar mean residual gradient, indexed effective orifice area and Doppler velocity index when adjusted to valve size. The frequency of mild (13.9% vs. 31.3%) and at least moderate (1.3% vs. 3.6%) PAR was less after Lotus than after SAPIEN3 implantation (P = 0.02). Multi-slice computed tomography-based annulus and left ventricular outflow tract diameters, calcification and percentage of oversizing were not different between those with or without mild or more PAR. On multivariate analysis, the use of Lotus valve was associated with less (odds ratio OR, 0.41, P = 0.03) occurrence of PAR. Conclusion Overall, haemodynamic performance was comparable between the Lotus and SAPIEN3 valves. Lotus valve required less oversizing and was associated with less PAR than SAPIEN3. |
| Starting Page | 157 |
| Ending Page | 167 |
| Page Count | 11 |
| File Format | PDF HTM / HTML |
| Alternate Webpage(s) | https://watermark.silverchair.com/jew280.pdf?token=AQECAHi208BE49Ooan9kkhW_Ercy7Dm3ZL_9Cf3qfKAc485ysgAAAbEwggGtBgkqhkiG9w0BBwagggGeMIIBmgIBADCCAZMGCSqGSIb3DQEHATAeBglghkgBZQMEAS4wEQQMKIOYZwea7hEC0YY8AgEQgIIBZNNwukMRZka082ANKuTePUlKa0xWXoTgKKHjFVFLBrZ4DUqZ66vRPcwcQAyyvaDU5VmHfoSA1ZX6qBGofB0enVcmI5VsR7DAgUI7Ss4__bG1W8x7G_d2yaNLoQ07-R0rW4nEpuTTcwr16Jg8nHVxR6tpMv1A85drLtTDmZ_fKcGG-eNvp2qzUPFeeJ9DGd7963LPsF9pZjTNVZmHSRKDxzd1T8bGPw1OXkKQxDZmvAEVbz5ZgrgH-zVYtHJJlijXa9k1rlkkXYqnjXuGH2ZHAaY2_P5bBc80CmK64SfRPHiwaBi9fiIY3XjOwc_FLC5UvP_zD7y-I5MSTOx4eluLPaPMOuEAOaIXAF3zCYD6aDHftNP9Oqge2TjB-NWcyv25R4NLf7zkxEoffvQStsPPvv-_oisU9BG-b4qHpoB5blpnlfcEu3HGxkGSWBpNnZ7CXA2fZhbykYSJs9pMLpYpBc_v6Jyn |
| PubMed reference number | 28064153v1 |
| Alternate Webpage(s) | https://doi.org/10.1093/ehjci/jew280 |
| DOI | 10.1093/ehjci/jew280 |
| Journal | European heart journal cardiovascular Imaging |
| Volume Number | 19 |
| Issue Number | 2 |
| Language | English |
| Access Restriction | Open |
| Subject Keyword | Anatomical orifice Aorta Aortic Valve Insufficiency Aortic Valve Stenosis Aortic valve structure Calcinosis Dilate procedure Echocardiography Eighty Exclusion Gradient Heart Valves Index Large Left Ventricular Outflow Tract Mitral Valve Prolapse Syndrome Outflow tract of left ventricle Patients Regurgitation Replacement of aortic valve (procedure) Seventy Nine Transcatheter Aortic Valve Replacement X-Ray Computed Tomography |
| Content Type | Text |
| Resource Type | Article |