Sensing failure associated with the Medtronic Sprint Fidelis defibrillator lead.
| Content Provider | Semantic Scholar |
|---|---|
| Author | Kenigsberg, David N. Mirchandani, Sunil Dover, Amanda N. Kowalski, Marcin Ryszard Wood, Mark A. Shepard, Richard K. Kalahasty, Gautham Stein, Kenneth Matthew Markowitz, Steven M. Iwai, Sei Shah, Bindi K. Lerman, Bruce B. Mittal, Suneet Ellenbogen, Kenneth A. |
| Copyright Year | 2008 |
| Abstract | INTRODUCTION The diameter of implantable cardioverter-defibrillator (ICD) leads has become progressively smaller over time. However, the long-term performance characteristics of these smaller ICD leads are unknown. METHODS We retrospectively evaluated 357 patients who underwent implantation of a Medtronic Sprint Fidelis defibrillating lead at two separate centers between September 2004 and October 2006. Lead characteristics were measured at implant, at early follow-up (1-4 days post implant), and every 3-6 months thereafter. RESULTS During the study period, 357 patients underwent implantation of the Medtronic Sprint Fidelis lead. The mean R-wave measured at implant through the device was not different (P = NS) when compared with that measured at first follow-up (10.5 +/- 5.0 mV vs 10.7 +/- 5.1 mV). Forty-one patients (13%) had an R-wave amplitude |
| File Format | PDF HTM / HTML |
| DOI | 10.1111/j.1540-8167.2007.01058.x |
| Alternate Webpage(s) | http://flahrs.com/wp-content/uploads/2016/09/Sensing-Failure-Associated-with-The-Medtronic-Sprint-Fidelis-Defibrillator-Lead..pdf |
| PubMed reference number | 18179527 |
| Alternate Webpage(s) | https://doi.org/10.1111/j.1540-8167.2007.01058.x |
| Journal | Medline |
| Volume Number | 19 |
| Issue Number | 3 |
| Journal | Journal of cardiovascular electrophysiology |
| Language | English |
| Access Restriction | Open |
| Content Type | Text |
| Resource Type | Article |