Usefulness of the Symphony Nitinol Stent for arteriosclerosis obliterans.
| Content Provider | Semantic Scholar |
|---|---|
| Author | Kinoshita, Yoshihisa Suzuki, Takahiko Hosokawa, Hiroaki Yokoya, Koichi |
| Copyright Year | 2002 |
| Abstract | The Symphony Peripheral Stent is a self-expanding stent made of thermal memory Nitinol wire. Stents were implanted in 39 lesions of 32 patients (26 men, 6 women) with atherosclerosis obliterans (ASO). The ankle-arm index (AAI), and vessel diameters evaluated by quantitative angiography were compared before and 6 months after treatment. Symphony Peripheral Stent implantation significantly improved the AAI from 0.50+/-0.4 to 0.9+/-0.2 (p<0.01), the minimum lumen diameter (MLD) from 2+/-1.5 to 5+/-1.4 mm (p<0.01) and percent diameter stenosis (% DS) from 69+/-20% to 16.5+/-8% (p<0.01). Re-evaluation of 33 of the 39 lesions 6 months after treatment revealed a low restenosis rate of 15%, an AAI of 0.8+/-0.3, MLD of 4.5+/-2 mm and %DS of 30+/-22%, so the Symphony Peripheral Stent is thus a promising choice for patients with ASO. |
| Starting Page | 299 |
| Ending Page | 308 |
| Page Count | 10 |
| File Format | PDF HTM / HTML |
| Alternate Webpage(s) | https://www.jstage.jst.go.jp/article/circj/66/11/66_11_1000/_pdf/-char/en |
| PubMed reference number | 12419929v1 |
| Volume Number | 66 |
| Issue Number | 11 |
| Journal | Circulation journal : official journal of the Japanese Circulation Society |
| Language | English |
| Access Restriction | Open |
| Subject Keyword | Ankle brachial pressure index (observable entity) Arteriosclerosis Obliterans Atherosclerosis Blood Vessel Tissue Diameter (qualifier value) Implants Inner Diameter Nitinol 55 Patients Peripheral Vascular Diseases Stenosis Stent Device Component Stent, device Structure of lumen of body system angiogram restenosis |
| Content Type | Text |
| Resource Type | Article |