Calibration of CR-39 solid state nuclear track detector
| Content Provider | Semantic Scholar |
|---|---|
| Author | Wanabongse, P. Sola, Banthom Jamsangtong, Jantanee Rattanabussayaporn, S. |
| Copyright Year | 2009 |
| Abstract | Nowadays there are increasing uses of SSNTD on scientific works in Thailand. Currently we are using SSNTD as an important tool for confirming active fault zones in a province in which a new nuclear research reactor has been proposed to come up. Soil gas radon was measured by both active and passive methods. In the latter case CR-39 SSNTD was installed in a PVC tube of 50 cm long by 5 cm in diameter. The tubes were placed in 50-cm deep holes lining perpendicular to the faults’ trace for one week. In this paper we describe the method and the results on calibrating these detectors using a radon chamber and a NIST-traceable radium-226 standard source. |