Loading...
Please wait, while we are loading the content...
Similar Documents
Quality Assurance (QA) of Volumetric Modulated Arc Therapy (VMAT) Treatment Plans: Analysis of 1000 Patients
| Content Provider | Semantic Scholar |
|---|---|
| Author | Daniel, Norma Largeron, G. Josserand, F. Daeira, E. Fafi, Samira Hajji, Zakia Wheeler, A. Khodri, Mustapha |
| Copyright Year | 2012 |
| Abstract | The analysis of the VMAT QA process performed over a clinical database of 1,000 patients was satisfactory, with results comparable with that of conventional IMRT. The EPID and IMSureQA® tools were proven to be suitable for routine clinical treatment plan QA. γ < 1 q Since May 2010, 1000 patients have benefited from the VMAT technique (VARIAN®). Patient specific QA was performed for each treatment: § Absolute dose measurements in water equivalent phantom. § Comparison of dose distributions calculated by Eclipse (AAA_10.0.28, grid 2mm, VARIAN®) with that measured by portal dosimetry using the EPIQA 2.1 (EPIDOS®) software. § Comparison of Eclipse® calculations with the IMSureQA® V3.3 (Standard Imaging® ) independent calculation algorithm. |
| File Format | PDF HTM / HTML |
| DOI | 10.1016/j.ijrobp.2012.07.2125 |
| Volume Number | 84 |
| Alternate Webpage(s) | http://astro2012.abstractsnet.com/handouts/011650_Quality_Assurance_Of_Volumetric_Modulated_Arc.pdf |
| Alternate Webpage(s) | https://doi.org/10.1016/j.ijrobp.2012.07.2125 |
| Language | English |
| Access Restriction | Open |
| Content Type | Text |
| Resource Type | Article |