Loading...
Please wait, while we are loading the content...
Similar Documents
Association of Inflow Cannula Position with Left Ventricular Unloading and Clinical Outcomes in Patients with HeartMate II Left Ventricular Assist Device
| Content Provider | Scilit |
|---|---|
| Author | Imamura, Teruhiko Nguyen, Ann Chung, Ben Rodgers, Daniel Sarswat, Nitasha Kim, Gene Raikhelkar, Jayant Adatya, Sirtaz Ota, Takeyoshi Song, Tae Juricek, Colleen Estep, Jerry D. Burkhoff, Daniel Jeevanandam, Valluvan Sayer, Gabriel Uriel, Nir |
| Copyright Year | 2019 |
| Description | Journal: Asaio journal The relationship between the HeartMate II left ventricular assist device (LVAD) position and pump thrombosis has been reported. However, further clinical implications of device position are unknown. This study aimed to investigate optimal device position for better left ventricular (LV) unloading and patient prognosis. Patients undergoing a ramp test with right heart catheterization after HeartMate II LVAD implantation were enrolled to this study. Device position was quantified from the chest X-ray obtained at the time of the ramp test: (1) inflow cannula angle relative to horizontal line, (2) pump angle relative to spine, (3) pump depth, (4) angle between inflow cannula and pump, and (5) angle between pump and outflow graft. LV unloading was assessed by pulmonary capillary wedge pressure at set LVAD speed. Fifty-four patients (60 years old and 34 male [63%]) were enrolled. Nobody experienced device malfunction during the study period. Increased LV unloading (i.e., lower pulmonary capillary wedge pressure) was associated with a narrower inflow cannula angle relative to horizontal line. Inflow cannula angle <75° was associated with higher 1 year heart failure readmission-free survival rate (p < 0.05, hazards ratio 7.56 [95% confidence interval 2.32–24.7]). In conclusion, HeartMate II LVAD inflow cannula position was associated with LV unloading and patient prognosis. Prospective studies to ensure optimal device positioning and target better clinical outcomes are warranted. |
| Related Links | http://links.lww.com/ASAIO/A290 https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6342671/pdf |
| Ending Page | 335 |
| Page Count | 5 |
| Starting Page | 331 |
| ISSN | 10582916 |
| e-ISSN | 1538943X |
| DOI | 10.1097/mat.0000000000000823 |
| Journal | Asaio journal |
| Issue Number | 4 |
| Volume Number | 65 |
| Language | English |
| Publisher | Ovid Technologies (Wolters Kluwer Health) |
| Publisher Date | 2019-05-01 |
| Access Restriction | Open |
| Subject Keyword | Journal: Asaio journal Cardiology and Cardiovascular Diseases |
| Content Type | Text |
| Resource Type | Article |
| Subject | Medicine Biomaterials Biophysics Bioengineering Biomedical Engineering |