Total ozone measurement: intercomparison of prototype new zealand filter instrument and dobson spectrophotometer
| Content Provider | NASA Technical Reports Server (NTRS) |
|---|---|
| Author | Basher, R. E. |
| Copyright Year | 1978 |
| Description | A five month intercomparison showed that the total ozone amounts of a prototype narrowband interference filter instrument were 7% less than those of a Dobson instrument for an ozone range of 0.300 to 0.500 atm cm and for airmasses less than two. The 7% bias was within the intercomparison calibration uncertainty. An airmass dependence in the Dobson instrument made the bias relationship airmass-dependent but the filter instrument's ozone values were generally constant to 2% up to an airmass of four. Long term drift in the bias was negligible. |
| File Size | 1649328 |
| Page Count | 18 |
| File Format | |
| Alternate Webpage(s) | http://archive.org/details/NASA_NTRS_Archive_19780020469 |
| Archival Resource Key | ark:/13960/t57d7pd94 |
| Language | English |
| Publisher Date | 1978-07-01 |
| Access Restriction | Open |
| Subject Keyword | Spectrophotometers Air Filters Calibrating Measuring Instruments Drift Instrumentation Instrument Errors Air Masses Ozonometry Ntrs Nasa Technical Reports ServerĀ (ntrs) Nasa Technical Reports Server Aerodynamics Aircraft Aerospace Engineering Aerospace Aeronautic Space Science |
| Content Type | Text |
| Resource Type | Technical Report |