Loading...
Please wait, while we are loading the content...
Similar Documents
Absolute calibration of a hydrogen discharge lamp in the vacuum ultraviolet
| Content Provider | NASA Technical Reports Server (NTRS) |
|---|---|
| Author | Nealy, J. E. |
| Copyright Year | 1975 |
| Description | A low-pressure hydrogen discharge lamp was calibrated for radiant intensity in the vacuum ultraviolet spectral region on an absolute basis and was employed as a laboratory standard source in spectrograph calibrations. This calibration was accomplished through the use of a standard photodiode detector obtained from the National Bureau of Standards together with onsite measurements of spectral properties of optical components used. The stability of the light source for use in the calibration of vacuum ultraviolet spectrographs and optical systems was investigated and found to be amenable to laboratory applications. The lamp was studied for a range of operating parameters; the results indicate that with appropriate peripheral instrumentation, the light source can be used as a secondary laboratory standard source when operated under preset controlled conditions. Absolute intensity measurements were recorded for the wavelengths 127.7, 158.0, 177.5, and 195.0 nm for a time period of over 1 month, and the measurements were found to be repeatable to within 11 percent. |
| File Size | 5843325 |
| Page Count | 35 |
| File Format | |
| Alternate Webpage(s) | http://archive.org/details/NASA_NTRS_Archive_19760007069 |
| Archival Resource Key | ark:/13960/t6256d692 |
| Language | English |
| Publisher Date | 1975-12-01 |
| Access Restriction | Open |
| Subject Keyword | Ground Support Systems And Facilities (space) Hydrogen Far Ultraviolet Radiation Calibrating Spectrometers Optical Equipment Photodiodes Gas Discharge Tubes Ntrs Nasa Technical Reports ServerĀ (ntrs) Nasa Technical Reports Server Aerodynamics Aircraft Aerospace Engineering Aerospace Aeronautic Space Science |
| Content Type | Text |
| Resource Type | Technical Report |