A Clinical Trial of the Accuracy and Treatment Experience of the Flash Glucose Monitor FreeStyle Libre in Adults with Type 1 Diabetes.
| Content Provider | Europe PMC |
|---|---|
| Author | Ólafsdóttir, Arndís F. Attvall, Stig Sandgren, Ulrika Dahlqvist, Sofia Pivodic, Aldina Skrtic, Stanko Theodorsson, Elvar Lind, Marcus |
| Copyright Year | 2017 |
| Abstract | AbstractBackground: In Sweden, FreeStyle Libre a flash glucose monitoring system came onto the market in 2014 as a complement to self-monitoring of blood glucose. The aim of this study was to evaluate the accuracy and treatment experience of the FreeStyle Libre system.Methods: Fifty-eight adults with type 1 diabetes used FreeStyle Libre for 10–14 days and measured capillary blood glucose levels with the HemoCue blood glucose measurement system at least six times a day simultaneously.Results: For the entire study period, the mean absolute relative difference (MARD) was 13.2% (95% confidence interval [CI] 12.0%–14.4%). MARD was 13.6% (95% CI 12.1%–15.4%) during week 1 and 12.7% (95% CI 11.5%–13.9%) during week 2. The mean absolute difference (MAD) for the whole study period was 19.8 mg/dL (1.1 mmol/L) (95% CI 17.8–21.8 mg/dL), including 20.5 mg/dL (1.14 mmol/L) during week 1 and 19.0 mg/dL (1.05 mmol/L) during week 2. The overall correlation coefficient was 0.96. For glucose values <72, 72–180, and >180 mg/dL (<4, 4–10, and >10 mmol/L), the MARD was 20.3% (95% CI 17.7%–23.1%), 14.7% (95% CI 13.4%–16%), and 9.6% (95% CI 8.5%–10.8%), respectively, and respective MAD values were 12.3, 17.8, and 23.6 mg/dL (0.69, 0.99, and 1.31 mmol/L). Using the 10-item visual analog scale, patients rated their experience with FreeStyle Libre as generally positive, with mean values ranging from 8.22 to 9.8.Conclusions: FreeStyle Libre had a similar overall MARD as continuous blood glucose monitoring systems in earlier studies when studied in similar at-home conditions. The overall patient satisfaction was high. |
| Related Links | https://europepmc.org/backend/ptpmcrender.fcgi?accid=PMC5359691&blobtype=pdf |
| Page Count | 9 |
| ISSN | 15209156 |
| Journal | Diabetes Technology & Therapeutics [Diabetes Technol Ther] |
| Volume Number | 19 |
| DOI | 10.1089/dia.2016.0392 |
| PubMed Central reference number | PMC5359691 |
| Issue Number | 3 |
| PubMed reference number | 28263665 |
| e-ISSN | 15578593 |
| Language | English |
| Publisher | Mary Ann Liebert, Inc. |
| Publisher Date | 2017-03-06 |
| Publisher Place | 140 Huguenot Street, 3rd FloorNew Rochelle, NY 10801USA |
| Access Restriction | Open |
| Rights License | This Open Access article is distributed under the terms of the Creative Commons Attribution Noncommercial License (http://creativecommons.org/licenses/by-nc/4.0/) which permits any noncommercial use, distribution, and reproduction in any medium, provided the original author(s) and the source are credited. © Arndís F. Ólafsdóttir et al., 2017; Published by Mary Ann Liebert, Inc. |
| Subject Keyword | Diabetes mellitus Flash glucose monitor Accuracy |
| Content Type | Text |
| Resource Type | Article |
| Subject | Endocrinology, Diabetes and Metabolism Medical Laboratory Technology Endocrinology |