Establishement Of The Magnetic Flux Density Standard In The Range 100 µT To 1000 µT
| Content Provider | CSIR - National Physical Laboratory (NPL) |
|---|---|
| Author | Aloysius, R. P. Verma, Vivek Pandey, Vibhav Srinivasan, P. Kotnala, R. K. |
| Abstract | The magnetic flux density standard in the range 100 (So(BT to 1000 (So(BT has been established using a standard Helmholtz coil. The ambient magnetic field including earth's magnetic field components inside the room and other noises has been compensated to a level of the order of 100 nT. A circular triaxial Helmholtz coil pair of size nearly 800 mm has been used for compensating the ambient magnetic field. The standard Helmholtz coil of size 300 mm has been placed in the geometrical centre of the triaxial Helmholtz coil pair. The magnetic flux density has been measured using triaxial and single axis fluxgate magnetometers. The coil constant of the standard Helmholtz coil has been found to be 850.6808� 0.0036804 (So(BT A, which is in good agreement with the manufacture's specification viz, 850 (So(BT A, with an uncertainly of ; � 0.5 %. |
| Related Links | http://npl.csircentral.net/140/1/113a.pdf http://dx.doi.org/10.1109/CPEM.2010.5543150 |
| File Format | |
| Journal | In: 2010 Conference on Precision Electromagnetic Measurements, 13-18 June 2010, Daejeon Convention Center, Daejeon |
| Language | English |
| Access Restriction | Open |
| Subject Keyword | Physics |
| Content Type | Text |
| Resource Type | Article |